LIFEVEST WCD 4000 SYSTEM
MDR report key: 3173440
·
Received June 12, 2013
Report
- Report Number
- 3008642652-2013-01535
- Event Type
- Malfunction
- Date Received
- June 12, 2013
- Date of Event
- May 8, 2013
- Report Date
- June 6, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DOES NOT RECOGNIZE BATTERY PACK) WAS CONFIRMED. UPON EVAL, THERE WAS CONTAMINATION ON THE BATTERY BOARD INSIDE THE CHARGER/MODEM. THE CAUSE OF THE INABILITY TO RECOGNIZE A BATTERY IS THE CONTAMINATION COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY LIQUID INGRESS OF AN UNK CONTAMINANT. NO ADVERSE EVENT RESULTED FROM THE CONTAMINATED CHARGER/ MODEM. THE PATIENT RECEIVED A REPLACEMENT CHARGER/MODEM.
Description of Event or Problem · 1
A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THIS CHARGER/MODEM WILL NOT RECOGNIZE A BATTERY PACK. THE PATIENT WAS ISSUED A REPLACEMENT CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 265921 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |