FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 3173379 · Received June 17, 2013

Report

Report Number
3005075853-2013-03075
Event Type
Malfunction
Date Received
June 17, 2013
Report Date
May 31, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE (A) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. DURING ANALYSIS THE JAWS OPEN AND CLOSE AS INTENDED. IN ADDITION THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS FOUND THAT THE DEVICE (B) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. DURING ANALYSIS THE JAWS OPEN AND CLOSE AS INTENDED. IN ADDITION THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. DEVICE B ADDITIONAL INFORMATION: BATCH # K90T0K, EXPIRATION DATE: 03/2018, MANUFACTURING DATE: 04/2013.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CLIPS WERE SCISSORING. A SECOND DEVICE WAS PULLED AND THE SAME THING OCCURRED. A THIRD DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273159 LIGAMAX-5MM ENDO CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1