FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC CLIP APPLIER

MDR report key: 3173348 · Received June 17, 2013

Report

Report Number
3005075853-2013-03073
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
May 29, 2013
Report Date
June 3, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE INSTRUMENT WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING THE DEVICE FOR FUNCTIONALITY THE DEVICE WAS CYCLED, FED, AND FORMED TWO CLIPS AS INTENDED AND IT EJECTED FIFTEEN CLIPS. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSECTOMY PROCEDURE, THE CLIP WAS NOT FED INTO THE JAWS PROPERLY AND THE CLIP FELL OUT FROM THE JAWS OUTSIDE OF THE PATIENT AT THE 2ND FIRING. THE CLIP DID NOT FALL INTO THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.