FDA Adverse Event
Malfunction
Summary report: N
REMB SAG SAW
MDR report key: 3173338
·
Received June 17, 2013
Report
- Report Number
- 0001811755-2013-01416
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- June 5, 2013
- Report Date
- June 5, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION, WEAR WAS FOUND ON THE BOOT AND THREE BEARINGS. ADDITIONALLY THE MOTOR WAS DAMAGED.
Description of Event or Problem · 1
THE REMB SAG SAW WAS RETURNED FOR SERVICE. UPON EVALUATION AT THE MANUFACTURER, IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. IT WAS ALSO REPORTED THAT DURING EVALUATION AT THE MANUFACTURER THAT THERE WAS A SUBSTANCE ON THE ROTOR WHICH COULD INDICATE THAT THE DEVICE WAS LEAKING. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES TO THE USER ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274511 | REMB SAG SAW | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |