FDA Adverse Event
Malfunction
Summary report: N
LOANER CORE SAG SAW
MDR report key: 3173334
·
Received June 17, 2013
Report
- Report Number
- 0001811755-2013-01411
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- June 4, 2013
- Report Date
- June 4, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION, THE FOLLOWING PARTS WERE WORN: THE BOOT, THE DRIVE BEARING AND THE ROTOR BEARING.
Description of Event or Problem · 1
THE CORE SAG SAW WAS RETURNED FOR SERVICE. UPON EVALUATION AT THE MANUFACTURER, IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. ADDITIONALLY, WHILE TESTING THE DEVICE DURING ROUTINE SERVICING, IT WAS REPORTED THAT THERE WAS A SUBSTANCE ON THE ROTOR ASSEMBLY WHICH COULD INDICATE THAT THE DEVICE WAS LEAKING. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES TO THE USER ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273051 | LOANER CORE SAG SAW | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |