FDA Adverse Event
Malfunction
Summary report: N
ACCU-CHEK ® MOBILE TEST STRIPS
MDR report key: 3173316
·
Received June 17, 2013
Report
- Report Number
- 1823260-2013-03632
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- May 18, 2013
- Report Date
- July 16, 2013
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- LFR
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
CUSTOMER IS IN HIS "MID-(B)(6)." THE EVENT OCCURRED IN (B)(6). WHILE THIS PRODUCT IS NOT SOLD IN THE UNITED STATES, IT IS LIKE OR SIMILAR TO A PRODUCT MARKETED IN THE UNITED STATES.
Description of Event or Problem · 1
REPORTER STATED THAT PATIENT RECEIVED THE FOLLOWING RESULTS ON PATIENT#S MOBILE SYSTEM 2 COMPARED TO A PROFESSIONAL METER WITHIN 10 MINUTES: 27 MMOL/L (MOBILE SYSTEM 2) AND 11.0 MMOL/L (PROFESSIONAL METER). NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT DUE TO THE DEVICE REPORTED. THE MANUFACTURER REQUESTED RETURN OF SUSPECT PRODUCT FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273046 | ACCU-CHEK ® MOBILE TEST STRIPS | BLOOD GLUCOSE MONITORING TEST STRIPS | LFR | ROCHE DIAGNOSTICS | NA | 278225 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |