FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3173305 · Received June 17, 2013

Report

Report Number
0001811755-2013-01409
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
June 4, 2013
Report Date
June 4, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION, CORROSION WAS FOUND ON THE MOTOR CARTRIDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CORE IMPACTION DRILL HEATED UP DURING A CASE. THERE WAS NO PATIENT OR USER INJURY REPORTED AND NO ADVERSE CONSEQUENCES ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272975 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1