FDA Adverse Event Malfunction Summary report: N

XMAX MOTOR

MDR report key: 3173257 · Received June 17, 2013

Report

Report Number
1045834-2013-02479
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
February 21, 2013
Report Date
February 22, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
ERL
PMA / PMN Number
K965080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE HAS BEEN RETURNED. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED. EVIDENCE INDICATED THIS WAS DUE TO USAGE WEAR OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PRETESTING THE HAND PIECE DEVICE WAS OVERHEATING AND WAS NOT ROTATING THE ATTACHMENT DEVICE. THE DEVICE WAS NOT USED IN A SURGICAL PROCEDURE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS ALLEGED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
274384 XMAX MOTOR DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HAND PIECE ERL DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention ATTACHMENT