FDA Adverse Event
Malfunction
Summary report: N
XMAX MOTOR
MDR report key: 3173257
·
Received June 17, 2013
Report
- Report Number
- 1045834-2013-02479
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- February 21, 2013
- Report Date
- February 22, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- K965080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE ACTUAL DEVICE HAS BEEN RETURNED. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE REPORTED CONDITION WAS DUPLICATED AND CONFIRMED. EVIDENCE INDICATED THIS WAS DUE TO USAGE WEAR OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING PRETESTING THE HAND PIECE DEVICE WAS OVERHEATING AND WAS NOT ROTATING THE ATTACHMENT DEVICE. THE DEVICE WAS NOT USED IN A SURGICAL PROCEDURE. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS ALLEGED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274384 | XMAX MOTOR | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HAND PIECE | ERL | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention | ATTACHMENT |