FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3173250 · Received June 17, 2013

Report

Report Number
0001811755-2013-01405
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
June 3, 2013
Report Date
June 3, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE MOTOR ASSEMBLY AND WEAR WAS FOUND ON THE SPINDLE HOUSING AND THE ROTOR COUPLER.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THE CORE IMPACTION DRILL HAD AN OVER TEMPERATURE WARNING ON THE CONSOLE. WHILE TESTING THE DEVICE AT THE MANUFACTURER IT WAS REPORTED THAT THE DEVICE HEATED UP. THERE WAS NO PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272939 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1