FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 3173250
·
Received June 17, 2013
Report
- Report Number
- 0001811755-2013-01405
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- June 3, 2013
- Report Date
- June 3, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION CORROSION WAS FOUND ON THE MOTOR ASSEMBLY AND WEAR WAS FOUND ON THE SPINDLE HOUSING AND THE ROTOR COUPLER.
Description of Event or Problem · 1
IT WAS REPORTED BY THE CUSTOMER THAT THE CORE IMPACTION DRILL HAD AN OVER TEMPERATURE WARNING ON THE CONSOLE. WHILE TESTING THE DEVICE AT THE MANUFACTURER IT WAS REPORTED THAT THE DEVICE HEATED UP. THERE WAS NO PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272939 | CORE IMPACTION DRILL | DRIVER, WIRE, AND BONE DRILL, MANUAL | DZJ | STRYKER INSTRUMENTS-KALAMAZOO |