FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3173240 · Received June 7, 2013

Report

Report Number
3008642652-2013-01523
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
May 10, 2013
Report Date
June 5, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION, THE CHARGER/MODEM WAS UNABLE TO POWER UP. THE CAUSE FOR THE POWER-UP FAILURE WAS ISOLATED TO A DAMAGED POWER SUPPLY CONNECTOR. THE ROOT CAUSE FOR THE DAMAGED POWER SUPPLY CONNECTOR COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CONNECTOR. THE LAST PATIENT TO USE THIS BATTERY CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, BATTERY CHARGER/MODEM SN (B)(4) WAS UNABLE TO POWER UP. THE LAST PATIENT TO USE THIS BATTERY CHARGER/MODEM DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253590 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA