FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3173238 · Received June 7, 2013

Report

Report Number
3008642652-2013-01526
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
February 6, 2013
Report Date
June 6, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER UP) WAS CONFIRMED. UPON INVESTIGATION, THE CHARGER/MODEM WAS UNABLE TO POWER UP. THE CAUSE FOR THE POWER-UP FAILURE WAS ISOLATED TO A DEFECTIVE PXA (U104) CHIP ON THE CHARGER C/A BOARD. THE ROOT CAUSE FOR THE DEFECTIVE U104 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE U104 PXA COMPONENT. THE PATIENT RECEIVED A REPLACEMENT BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A (B)(6) DISTRIBUTOR CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT A PATIENT'S BATTERY CHARGER/MODEM WAS NOT POWERING UP. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253561 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA