FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3173232
·
Received June 7, 2013
Report
- Report Number
- 3008642652-2013-01488
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- May 13, 2013
- Report Date
- May 31, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY: DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. UPON EVAL, THE MONITOR'S REAR RESPONSE BUTTON WAS NOT FUNCTIONAL. THE ROOT CAUSE FOR THE FAILURE WAS UNABLE TO BE POSITIVELY DETERMINED BUT WAS LIKELY A DEFECTIVE RESPONSE BUTTON SWITCH. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE RESPONSE BUTTON. THE LAST PT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, THE REAR RESPONSE BUTTON ON MONITOR SN (B)(4) WAS NOT WORKING. THE LAST PT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.