FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3173194 · Received May 31, 2013

Report

Report Number
3008642652-2013-01465
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
April 26, 2013
Report Date
May 29, 2013
Manufacturer
ZOLL LIFECOR CORP
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK BELT MESSAGES) HAS BEEN CONFIRMED. UPON EVALUATION, THE BROWN (-5V) WIRE WAS OPEN IN THE TRUNK CABLE CONNECTOR. THE CAUSE OF THE CHECK BELT MESSAGES IS THE OPEN BROWN WIRE. THE ROOT CAUSE OF THE OPEN WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE AT THE CONNECTOR. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK BELT MESSAGES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
242413 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORP WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 30 YR