FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3173194
·
Received May 31, 2013
Report
- Report Number
- 3008642652-2013-01465
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Date of Event
- April 26, 2013
- Report Date
- May 29, 2013
- Manufacturer
- ZOLL LIFECOR CORP
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK BELT MESSAGES) HAS BEEN CONFIRMED. UPON EVALUATION, THE BROWN (-5V) WIRE WAS OPEN IN THE TRUNK CABLE CONNECTOR. THE CAUSE OF THE CHECK BELT MESSAGES IS THE OPEN BROWN WIRE. THE ROOT CAUSE OF THE OPEN WIRE CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE AT THE CONNECTOR. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A (B)(6) FEMALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK BELT MESSAGES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.