LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-01438
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Date of Event
- May 8, 2013
- Report Date
- May 24, 2013
- Manufacturer
- ZOLL LIFECOR CORP
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER ON PAST MAIN SCREEN) WAS CONFIRMED. UPON INVESTIGATION THE MONITOR WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO AN INTERMITTENT BGA SOLDER CONNECTION AT THE U500 DSP COMPONENT ON THE MONITOR C/A BOARD. THE ROOT CAUSE FOR THE INTERMITTENT BGA CONNECTION AT U500 COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR IMPACTING A HARD SURFACE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE ((B)(4)) TO REDUCE THE PCA STRAIN WAS APPROVED BY FDA ON (B)(4) 2012. IMPLEMENTATION BEGAN (B)(4) 2013. RESULTS WILL BE MONITORED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE U500 COMPONENT. THE PT RECEIVED A REPLACEMENT MONITOR.
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PT'S MONITOR WAS NOT POWERING UP PROPERLY. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242412 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORP | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR |