FDA Adverse Event Malfunction Summary report: N

INSTINCT ENDOSCOPIC HEMOCLIP

MDR report key: 3173166 · Received June 5, 2013

Report

Report Number
1037905-2013-00359
Event Type
Malfunction
Date Received
June 5, 2013
Report Date
May 7, 2013
Manufacturer
WILSON-COOK MEDICAL INC.
Product Code
MND
PMA / PMN Number
K121505
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: OUR LABORATORY EVALUATION OF THE PRODUCT SAID TO BE INVOLVED FOUND THE DRIVE WIRE SEPARATED INSIDE THE HANDLE. THE PROXIMAL END OF THE DRIVE WIRE WAS BOWED INDICATING PROPER ASSEMBLY OF THE SECURING COMPONENT. USING A HEMOSTAT TO GRASP THE DRIVE WIRE, THE CLIP WAS OPENED AND CLOSED. A SLIGHT INCREASE IN RESISTANCE WAS OBSERVED IN THE MOVEMENT OF THE DRIVE WIRE ONCE THE CLIP WAS PULLED HALFWAY INTO THE HOUSING PRIOR TO DEPLOYMENT. THE DEVICE WAS ADVANCED INTO OLYMPUS GIF Q20 2.8 ENDOSCOPE IN A SIMULATED UPPER GI POSITION WITH THE TIP IN THE MAXIMUM RETROFLEXED POSITION TO SIMULATE A WORST CASE POSITION. USING THE HEMOSTATS, THE CLIP DID SUCCESSFULLY DEPLOY ON SIMULATED TISSUE. THE DRIVE WIRE WAS REMOVED FROM THE DEVICE, VISUALLY EXAMINED, AND DETERMINED THAT THE DRIVE WIRE WAS MANUFACTURED CORRECTLY. THE HOOK OF THE DRIVE WIRE IS INTACT. A PRODUCT-SPECIFIC DISCREPANCY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THIS OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: RESISTANCE ENCOUNTERED WHILE ATTEMPTING TO DEPLOY THE CLIP AFTER THE CLIP IS CLOSED ON THE TISSUE (I.E. DIFFICULTY WITH CLIP RELEASE) CAN LEAD TO DAMAGE OF DEVICE COMPONENTS SUCH AS THE HOOK AT THE DISTAL END OF THE DRIVE WIRE OR THE SECURITY OF THE DRIVE WIRE TO THE HANDLE. THE INSTRUCTION FOR USE INSTRUCTS THE USER TO VERIFY SMOOTH HANDLE OPERATION AND CLIP OPERATION PRIOR TO USE. PRIOR TO DISTRIBUTION, ALL INSTINCT ENDOSCOPIC HEMOCLIPS ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TESTING TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A CORRECTIVE ACTION HAS BEEN INITIATED IN AN EFFORT TO REDUCE OCCURRENCES OF THIS NATURE. THIS PRODUCT WAS MANUFACTURED PRIOR TO IMPLEMENTATION OF THIS CORRECTIVE ACTION. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 1

DURING AN ENDOSCOPY PROCEDURE, A COOK INSTINCT ENDOSCOPIC HEMOCLIP WAS USED. THE CLIP WAS ATTACHED TO THE TISSUE SITE. THE CLIP WOULD NOT RELEASE FROM THE DEPLOYMENT DEVICE. WHILE APPLYING PRESSURE, THE INTERNAL DRIVE WIRE DISCONNECTED FROM THE HANDLE. THE CLIP WAS PULLED OFF THE TISSUE SITE WHILE IN A MOSTLY CLOSED BUT SLIGHTLY OPEN POSITION. ANOTHER DEVICE OF THE SAME TYPE WAS USED TO FINISH THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249402 INSTINCT ENDOSCOPIC HEMOCLIP MND, LIGATOR, ESOPHAGEAL MND WILSON-COOK MEDICAL INC. W3255798

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS ENDOSCOPE: MODEL #UNK