FDA Adverse Event Malfunction Summary report: N

FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME

MDR report key: 3173158 · Received June 5, 2013

Report

Report Number
1037905-2013-00361
Event Type
Malfunction
Date Received
June 5, 2013
Date of Event
May 3, 2013
Report Date
May 7, 2013
Manufacturer
WILSON-COOK MEDICAL INC.
Product Code
KNS
PMA / PMN Number
K052051
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVALUATION. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVALUATION. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. IMPROPER CUTTING WIRE ORIENTATION CAN OCCUR IF THE DISTAL END OF THE CATHETER IS SHAPED MANUALLY. THIS SPHINCTEROTOME CATHETER IS PRECURVED AND IS PROVIDED WITH A PRECURVED STYLET IN THE DISTAL TIP OF THE CATHETER. THIS OBVIATES THE NEED FOR MANUAL FORMATION. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT APPLY MANUAL PRESSURE TO TIP OR CUTTING WIRE OF THE SPHINCTEROTOME TO INFLUENCE ORIENTATION, AS THIS MAY RESULT IN DAMAGE TO DEVICE." OTHER FACTORS THAT CAN CONTRIBUTE TO IMPROPER CUTTING WIRE ORIENTATION INCLUDE MANIPULATING THE HANDLE WITH THE CATHETER IN A COILED POSITION OR WITH THE PRECURVED STYLET INSIDE THE CANNULATING TIP. THE INSTRUCTIONS FOR USE ADVISE THE USER TO UNCOIL AND STRAIGHTEN THE SPHINCTEROTOME UPON REMOVING THE DEVICE FROM THE PACKAGING. THE USER IS THEN INSTRUCTED TO CAREFULLY REMOVE THE PRECURVED STYLET FROM THE CANNULATING TIP. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING COMMENT: "NOTE: DO NOT EXERCISE HANDLE WHILE DEVICE IS COILED OR PRECURVED STYLET IS IN PLACE, AS THIS MAY CAUSE DAMAGE TO SPHINCTEROTOME AND RENDER IT INOPERABLE." PRIOR TO DISTRIBUTION, ALL FUSION OMNI-TOME SPHINCTEROTOMES ARE SUBJECTED TO A VISUAL INSPECTION AND FUNCTIONAL TEST TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ENDOSCOPIC PROCEDURE, A COOK FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME WAS USED. THE ORIENTATION OF THE SPHINCTEROTOME WAS INCORRECT. ANOTHER OF THE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249776 FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) KNS WILSON-COOK MEDICAL INC. W3236189

Patients

Seq Age Sex Outcome Treatment
1 PENTAX ED 3485T ENDOSCOPE| COOK METII 35-260 WIRE GUIDE