FDA Adverse Event Malfunction Summary report: N

FENESTRATED BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3173134 · Received June 17, 2013

Report

Report Number
2955842-2013-02188
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
May 6, 2013
Report Date
May 28, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE ONE GRIP WAS BENT, CAUSING SIDE-TO-SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A .17 OFFSET AT THE TIPS. THE BENT GRIP HAD SEPARATION OF THE YAW PULLEY AND THE PULLEY COVER AT THE GLUE JOINT, INDICATING OVERLOADING AT THE TIP. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY DUE TO MISHANDLING. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. ENGINEERING ALSO OBSERVED THAT THE INSTRUMENT PITCH CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. NO DAMAGE WAS FOUND AT THE CLEVIS. THE FRAYED SEGMENT WAS APPROXIMATELY 0.03 IN LENGTH. NO OTHER CABLE DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS O HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. O DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING EVALUATION FOUND THAT THE ONE GRIP WAS BENT, CAUSING SIDE-TO-SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A .17 OFFSET AT THE TIPS. THE BENT GRIP HAD SEPARATION OF THE YAW PULLEY AND THE PULLEY COVER AT THE GLUE JOINT, INDICATING OVERLOADING AT THE TIP. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY DUE TO MISHANDLING. THE INSTRUMENT PASSED ELECTRICAL CONTINUITY TESTING. ENGINEERING ALSO FOUND THE PITCH CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. NO DAMAGE WAS FOUND AT THE CLEVIS. THE FRAYED SEGMENT WAS APPROXIMATELY 0.03 IN LENGTH. NO OTHER CABLE DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT SHAFT WAS NOTED TO BE BENT AND WOULD NOT GRASP OR CLOSE. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
274078 FENESTRATED BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120629 132

Patients

Seq Age Sex Outcome Treatment
1 DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES