LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-01470
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Date of Event
- May 9, 2013
- Report Date
- May 29, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION THE MONITOR FAILED A PULSE TEST DUE TO RESISTOR R45 BEING OPEN ON THE C/A BOARD. THE CAUSE FOR THE OPEN R45 RESISTOR WAS A BROKEN LEAD ON HIGH-VOLTAGE CAPACITOR C21 ON THE DEFIBRILLATOR BOARD. THE ROOT CAUSE FOR THE BROKEN LEAD ON C21 COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR IMPACTING A HARD SURFACE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE TO REDUCT THE PCA STRAIN WAS APPROVED BY FDA ON (B)(4) 2012. IMPLEMENTATION BEGAN (B)(4) 2013. RESULTS WILL BE MONITORED. NO ADVERSE EVENT RESULTED FROM THE BROKEN LEAD ON C21. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, MONITOR SN (B)(4) FAILED A PULSE TEST. THE LAST PATIENT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 241882 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |