FDA Adverse Event Malfunction Summary report: N

ECHELON*FLEX60

MDR report key: 3173072 · Received June 17, 2013

Report

Report Number
3005075853-2013-03072
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
May 23, 2013
Report Date
May 23, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K081146
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. HOW WAS THE PATIENT IMPACTED? WAS THE DEVICE FIRED ON TISSUE THAT HAD BEEN RADIATED OR HAS THE PATIENT BEEN TAKING SYSTEMIC STEROIDS? ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 12TH, ETC)? IF ECHELON, DURING WHICH STROKE DID THE EVENT OCCUR? WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)? AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? IS THERE ANY OTHER ADDITIONAL INFORMATION YOU WOULD LIKE TO SHARE ABOUT THIS DEVICE OR PROCEDURE?

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH ONE ECR60G CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED FULLY FIRED AND IN GOOD VISUAL CONDITION. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DIFFICULTIES NOTED DURING THE FUNCTIONAL TESTING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A THORACIC PROCEDURE, ON THE FOURTH FIRING USING A GREEN RELOAD; THE BLADE TORE THROUGH THE BRONCHUS. THE STAPLE LINE WAS INCOMPLETE; SOME STAPLES WERE STILL IN THE RELOAD AFTER FIRING. THE CASE IS STILL IN SESSION AND THE SURGEON WILL RE-CUT AND STAPLE THE BRONCHUS USING A SECOND DEVICE. PREVIOUS CARTRIDGES USED PRIOR TO THE EVENT IS UNKNOWN. IT IS UNKNOWN IF BUTTRESSING MATERIAL WAS USED. THERE WERE NO ISSUES WITH THE REMOVAL OF THE DEVICE FROM TISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273708 ECHELON*FLEX60 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA K4CJ16

Patients

Seq Age Sex Outcome Treatment
1