ECHELON*FLEX60
Report
- Report Number
- 3005075853-2013-03072
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- May 23, 2013
- Report Date
- May 23, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. HOW WAS THE PATIENT IMPACTED? WAS THE DEVICE FIRED ON TISSUE THAT HAD BEEN RADIATED OR HAS THE PATIENT BEEN TAKING SYSTEMIC STEROIDS? ON WHICH FIRING(S) DID THIS EVENT OCCUR (1ST, 2ND, 12TH, ETC)? IF ECHELON, DURING WHICH STROKE DID THE EVENT OCCUR? WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)? AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? IS THERE ANY OTHER ADDITIONAL INFORMATION YOU WOULD LIKE TO SHARE ABOUT THIS DEVICE OR PROCEDURE?
(B)(4). THE ANALYSIS FOUND THAT ONE EC60A DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH ONE ECR60G CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED FULLY FIRED AND IN GOOD VISUAL CONDITION. THE DEVICE WAS TESTED FOR FUNCTIONALITY IN THE ARTICULATED POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE PERFORMED WITHOUT ANY DIFFICULTIES NOTED DURING THE FUNCTIONAL TESTING. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A THORACIC PROCEDURE, ON THE FOURTH FIRING USING A GREEN RELOAD; THE BLADE TORE THROUGH THE BRONCHUS. THE STAPLE LINE WAS INCOMPLETE; SOME STAPLES WERE STILL IN THE RELOAD AFTER FIRING. THE CASE IS STILL IN SESSION AND THE SURGEON WILL RE-CUT AND STAPLE THE BRONCHUS USING A SECOND DEVICE. PREVIOUS CARTRIDGES USED PRIOR TO THE EVENT IS UNKNOWN. IT IS UNKNOWN IF BUTTRESSING MATERIAL WAS USED. THERE WERE NO ISSUES WITH THE REMOVAL OF THE DEVICE FROM TISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273708 | ECHELON*FLEX60 | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | K4CJ16 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |