FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3172930 · Received June 17, 2013

Report

Report Number
1416980-2013-15562
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
May 25, 2013
Report Date
May 25, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT AVAILABLE FOR ANALYSIS AND THE LOT NUMBER WAS UNKNOWN; THEREFORE A BATCH REVIEW COULD NOT BE PERFORMED. PER "THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE", USERS ARE INSTRUCTED THAT CLAMPS ON ANY UNUSED SOLUTION LINES MUST REMAIN CLOSED WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HOME PATIENT (HP) HAD RECEIVED A SYSTEM ERROR (SE) 2240 ALARM (AIR IN LINE), WHICH OCCURRED ON THE HOMECHOICE (HC) MACHINE DURING THE THERAPY, PATIENT WAS CONNECTED. THERE WAS AN OPEN CLAMP ON AN UNUSED SUPPLY LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) INFORMED THE HP TO START OVER WITH ALL NEW SUPPLIES OR PERFORM MANUAL THERAPY. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE, A FOLLOW UP MEDWATCH WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272907 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 50 YR HOMECHOICE