FDA Adverse Event Malfunction Summary report: N

LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR

MDR report key: 3172817 · Received June 17, 2013

Report

Report Number
3015876-2013-00510
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
May 20, 2013
Report Date
May 21, 2013
Manufacturer
PHYSIO-CONTROL, INC
Product Code
LDD
PMA / PMN Number
K063119
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE DEVICE WAS NOT EVALUATED BY PHYSIO-CONTROL. A THIRD PARTY AGENT EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. THE THIRD PARTY AGENT REPLACED THE DEVICE POWER SUPPLY ASSEMBLY. AFTER OBSERVING PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING, THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE DID NOT POWER ON EITHER ON AC MAIN POWER OR ON BATTERY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.