LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-01440
- Event Type
- Malfunction
- Date Received
- May 31, 2013
- Date of Event
- April 29, 2013
- Report Date
- May 24, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLES) WAS CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED A HI-POT AND INSULATION RESISTANCE TEST. THE FAILURE WAS CAUSED BY A DAMAGED CABLE. THE CABLE CONNECTING THE DISTRIBUTION NODE (DN) TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING THE PULSE WIRE. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.
A (B)(6) FEMALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HER ELECTRODE BELT CABLES WERE DAMAGED. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242056 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |