FDA Adverse Event Injury Summary report: N

XCEL DILATING TIP TROCAR

MDR report key: 3172758 · Received June 17, 2013

Report

Report Number
3005075853-2013-03060
Event Type
Injury
Date Received
June 17, 2013
Date of Event
May 22, 2013
Report Date
May 28, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GCJ
PMA / PMN Number
K032676
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS OF THE TROCAR PERFORMED FOUND THAT THE TROCAR WAS RECEIVED WITH THE SLEEVE CANNULA BROKEN. HIGHER MAGNIFICATION VIEW OF THE FRACTURE, NOTE THE DUCTILITY IN THE SURFACE INDICATING THE PART BROKE IN A DUCTILE MODE. THE MATERIAL APPEARS TO HAVE BEEN MOLDED PROPERLY AND THE CANNULA BROKE FROM AN EXCESSIVE SIDE LOAD APPLIED TO THE DEVICE. WE DID NOT RECEIVE A VALID BATCH OR LOT NUMBER FOR THE PRODUCT INVOLVED IN THIS COMPLAINT. THEREFORE, WE WERE UNABLE TO CHECK MANUFACTURING RECORDS FOR ANY RELATED NON-CONFORMANCES.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION WAS NOT PROVIDED BY THE INITIAL CONTACT. INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. REP PROVIDED THE FOLLOWING INFORMATION AFTER HE SPOKE TO THE SURGEON: THE SURGEON EXPLAINED THERE WAS NOT AN INSTRUMENT IN THE DEVICE WHEN THE BREAK OCCURRED. THE TROCAR HAD POPPED OUT AND SHE WAS RE-INSERTING IT WHEN IT BROKE. SHE STATED SHE WAS NOT PLACING EXCESSIVE FORCE ON THE TROCAR. THE PATIENT RECOVERED WELL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A GYN PROCEDURE THE SURGEON HAD PLACED THE PORT INTO THE PATIENT; IT KEPT COMING OUT OF THE TISSUE DUE TO THE PATIENT WAS MORBIDLY OBESE. THEY COULD NOT KEEP THE TROCAR IN THE BELLY OF THE PATIENT DUE TO EXCESS FATTY TISSUE IN THE FASCIA. THE SURGEON TRIED TO PLACE THE CAMERA INTO THE PORT WITHOUT THE OBTURATOR PRESENT DUE TO THE FATTY TISSUE PUSHING UP ON THE DEVICE. THE PORT KEPT POPPING OUT AND THEN THE CANNULA BROKE OFF IN THE SECTION ABOUT THREE RIBS UP ON THE CANNULA FROM THE TOP PORTION OF THE DEVICE. THEY HAD TO CALL IN A GENERAL SURGEON IN AT THAT POINT. THEY CONVERTED TO AN OPEN PROCEDURE. THEY THEN REMOVED THE PORT AND THE BROKEN TROCAR FROM THE PATIENT. THIS PROLONGED THE PROCEDURE OVER AN HOUR. THE PATIENT WAS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272929 XCEL DILATING TIP TROCAR LAPAROSCOPE, GENERAL AND PLASTIC SURGERY GCJ ETHICON ENDO-SURGERY, LLC. NA ASKU

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention