FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3100 SYSTEM

MDR report key: 3172756 · Received May 31, 2013

Report

Report Number
3008642652-2013-01481
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 8, 2013
Report Date
May 30, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (RESETTING) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO AN INTERMITTENT U200 PLD (PROGRAMMABLE LOGIC DEVICE) ON THE MONITOR C/A BOARD. THE ROOT CAUSE FOR THE INTERMITTENT U200 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE INTERMITTENT U200 COMPONENT. THE PATIENT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WAS RESETTING. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.