FDA Adverse Event
Malfunction
Summary report: N
MEDFUSION 3500 SYRINGE INFUSION PUMP
MDR report key: 3172748
·
Received June 3, 2013
Report
- Report Number
- 2183502-2013-00284
- Event Type
- Malfunction
- Date Received
- June 3, 2013
- Date of Event
- April 30, 2013
- Report Date
- May 31, 2013
- Manufacturer
- SMITHS MEDICAL MD
- Product Code
- FRN
- PMA / PMN Number
- K040899
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL: THE SUSPECT DEVICE WAS RETURNED AND EVALUATED. FUNCTIONAL TESTING CONFIRMED A CHECK CLUTCH ALARM DUE TO A LOOSE CARRIAGE WASHER; THEREFORE, THE CARRIAGE WASHER WAS TIGHTENED. FOLLOWING REPAIR, THE DEVICE PASSED ALL FUNCTION AND DELIVERY TESTING.
Description of Event or Problem · 1
A REPORT WAS RECEIVED THAT STATED THE PUMP ALARMED "CHECK CLUTCH." NO PT INJURY OR ADVERSE EVENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 244273 | MEDFUSION 3500 SYRINGE INFUSION PUMP | SYRINGE INFUSION PUMP | FRN | SMITHS MEDICAL MD | 3500 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |