FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3172732 · Received June 17, 2013

Report

Report Number
1416980-2013-15543
Event Type
Malfunction
Date Received
June 17, 2013
Date of Event
May 29, 2013
Report Date
May 29, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HOME PATIENT (HP) HAD A SYSTEM ERROR 2240 (AIR IN TUBING) AND SYSTEM ERROR 2367 ON THE HOMECHOICE (HC) DURING PREVIOUS THERAPY WHILE THE HP WAS CONNECTED. THE HP HAD DISCONNECTED FROM THE HC AND DISPOSED OF ALL THE SUPPLIES. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) REVIEWED THE ALARM LOG AND EXPLAINED THE POSSIBLE CAUSES OF THE ALARMS. THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
273683 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 59 YR HOMECHOICE