HOMECHOICE
Report
- Report Number
- 1416980-2013-15529
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- February 15, 2013
- Report Date
- May 23, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS USE ERROR, TIDAL TOTAL UF REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER?S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. PG 12-30 HAS THE WARNING THAT "A TOTAL UF VOLUME SET TOO LOW CAN RESULT IN A GRADUAL BUILDUP OF UF VOLUME DURING THE THERAPY. THIS CAN RESULT IN AN IIPV SITUATION. " SECTION 13 "TOPIC 4: TIDAL THERAPY" GIVES THE TRAINER INSTRUCTIONS ON HOW TO PROPERLY SET TIDAL THERAPY VALUES. ON PG 13-9 THE INSTRUCTIONS STATE "A GOOD STARTING POINT WOULD BE, AT A MINIMUM, TO REVIEW A DRAIN HISTORY OVER THE PREVIOUS SEVEN TREATMENT DAYS. THE TOTAL AMOUNT OF UF OVER THE SEVEN DAYS SHOULD BE ADDED AND THEN DIVIDED BY SEVEN TO OBTAIN AN AVERAGE DAILY UF GOAL." THE SUGGESTED SETTING FOR TOTAL UF IS DESCRIBED ON PG 12-30 AS "SEVENTY PERCENT OF THE NORMAL NIGHT UF IS A GOOD STARTING POINT FOR DETERMINING THE OPTIMUM TOTAL UF. " IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING EVALUATION OF A RETURNED HOMECHOICE MACHINE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2013 16:59:23. DURING NIGHT DRAIN CYCLE FOUR, THE PATIENT'S ULTRAFILTRATION READING WAS 2151ML, INDICATING THE HOME PATIENT (HP) DRAINED 1351ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 274152 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |