FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3172643 · Received May 31, 2013

Report

Report Number
3008642652-2013-01433
Event Type
Malfunction
Date Received
May 31, 2013
Date of Event
May 7, 2013
Report Date
May 24, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE OR WIRES EXPOSED) WAS CONFIRMED. UPON INVESTIGATION, THE DISTRIBUTION NODE (DN) TO REAR THERAPY ELECTRODE (TE) CABLE WAS PULLED FROM THE STRAIN RELIEF. THIS RESULTED IN AN INTERMITTENT CONNECTION BETWEEN THE CABLE AND DN. THE ROOT CAUSE OF THE DAMAGED CABLE COULD NOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE WIRING CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES HAD COME OUT. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
241465 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 63 YR