GYNECARE MORCELLEX* TISSUE MORCELLATOR
Report
- Report Number
- 2210968-2013-07332
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Date of Event
- April 11, 2013
- Report Date
- May 29, 2013
- Manufacturer
- ETHICON, INC.
- Product Code
- HET
- PMA / PMN Number
- K100280
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
(B)(4): THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE BLADE FAILED TO ROTATE DURING THE EVALUATION; IT IS LIKELY THAT THE ACCUMULATION OF BLOOD, BODY FLUIDS, AND TISSUE PREVENTED THE DEVICE FROM CONTINUING TO OPERATE AS INTENDED. SINCE THE BLADE WAS JAMMED DUE TO ACCUMULATION OF TISSUE, DEVICE WAS MAKING EXCESSIVE NOISE DURING EVALUATION.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY ON (B)(6) 2013. DURING THE PROCEDURE, THE DEVICE WAS MAKING A "GRINGING" NOISE AND CUTTING VERY SLOW. THE DEVICE CONTINUOUSLY WORKED, BUT VERY SLOW. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273319 | GYNECARE MORCELLEX* TISSUE MORCELLATOR | LAPAROSCOPE, GYNECOLOGIC | HET | ETHICON, INC. | NA | MT217502 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |