ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-08284
- Event Type
- Malfunction
- Date Received
- June 17, 2013
- Report Date
- May 20, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. THE PUMP BOOTED TO THE VERIFY SCREEN WITH DIM PINK CONTRAST. THE CONTRAST, UP, AND OK KEYS WERE FOUND TO BE WORN. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE DISPLAY WAS FOUND TO BE DIM/PINK. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 273296 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |