FDA Adverse Event Malfunction Summary report: N

SYMBIQ SINGLE CHANNEL

MDR report key: 3172389 · Received June 5, 2013

Report

Report Number
9615050-2013-01528
Event Type
Malfunction
Date Received
June 5, 2013
Date of Event
May 1, 2013
Report Date
May 10, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE INITIALLY PASSED TESTING; HOWEVER, A WHITESCREEN ERROR WAS DISPLAYED DURING FURTHER TESTING. TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT PASS THE SDRAM TEST WHICH MAY HAVE CAUSED THE CUSTOMER'S REPORTED WHITESCREEN ERROR. THIS IS DUE TO THE SOM2 HARDWARE MODULE. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. CUSTOMER NOTIFICATION DATED (B)(4) 2013. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED TO WHITESCREEN ERROR. THE PUMP WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "WHITE SCREEN." NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFICATION PT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS, AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY THE DEVICE WAS POWERED ON, THE DEVICE ALARMED WITH A WHITESCREEN ERROR. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249556 SYMBIQ SINGLE CHANNEL 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK