FDA Adverse Event Malfunction Summary report: N

GEMSTAR 7 THERAPY

MDR report key: 3172300 · Received June 5, 2013

Report

Report Number
9615050-2013-01512
Event Type
Malfunction
Date Received
June 5, 2013
Date of Event
May 6, 2013
Report Date
May 6, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K083019
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME THE CUSTOMER WILL NOT BE RETURNING THE DEVICE FOR EVALUATION. IF THE DEVICE IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE DEVICE WAS NOT RETURNED TO HOSPIRA FOR TESTING AND INVESTIGATION; THEREFORE, ATTRIBUTION OF THE ISSUE TO THE DEVICE COULD NOT BE DETERMINED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT ALARM WHEN A PROXIMAL OCCLUSION WAS PRESENT. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249458 GEMSTAR 7 THERAPY 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA