FDA Adverse Event Malfunction Summary report: N

ZIMMER SKIN GRAFT MESHER

MDR report key: 3172278 · Received June 5, 2013

Report

Report Number
1526350-2013-00293
Event Type
Malfunction
Date Received
June 5, 2013
Date of Event
May 1, 2013
Report Date
May 17, 2013
Manufacturer
ZIMMER SURGICAL
Product Code
GFD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR REPAIR AND EVAL. THE SERVICE RECORD INDICATES THAT THE DEVICE WAS MANUFACTURED ON 2/21/2002 AND WAS LAST REPAIRED ON 9/6/2012 FOR A NON-RELATED ISSUE. EVAL OF THE DEVICE OBSERVED THAT THE RATCHET GEAR WAS STUCK ON THE ROLLER. IT WAS ALSO OBSERVED THAT THE ROLLER, COMB, SIDE PLATES AND GEAR WERE DAMAGED. A TEST MEAN AND CALIBRATION COULD NOT BE PERFORMED DUE TO DAMAGE OF THE COMB. THE COMPLAINT WAS LIKELY CAUSED BY ROLLER AND RATCHET GEAR DAMAGED DUE TO IMPROPER HANDLING BY THE CUSTOMER. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ZIMMER SKIN GRAFT MESHER WAS BROKEN. DURING DEVICE ANALYSIS, IT WAS DETERMINED THAT THE COMB WAS BENT. CLINICAL FOLLOW UP WITH THE CUSTOMER COULD NOT PROVIDE ANY ADDITIONAL INFO REGARDING THE REPORTED EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
249371 ZIMMER SKIN GRAFT MESHER ZIMMER SKIN GRAFT MESHER GFD ZIMMER SURGICAL NA NA

Patients

Seq Age Sex Outcome Treatment
1 ZIMMER SKIN GRAFT MESHER CUTTERS:| 00-7703-020-00, SN# (B)(4)| 00-7703-030-00, SN# (B)(4)