FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3172149 · Received June 11, 2013

Report

Report Number
1720753-2013-06967
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 29, 2013
Report Date
June 11, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE SYSTEM INTERFACE BOARD. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THERE WAS A LOSS OF LIVE IMAGE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264516 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1