FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3172142 · Received June 11, 2013

Report

Report Number
1720753-2013-06975
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 31, 2013
Report Date
June 11, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR WAS CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A FILAMENT REGULATOR FAILURE MESSAGE DURING A PROCEDURE WITH PATIENT INVOLVEMENT, THEREFORE THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ARO. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
263442 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1