FDA Adverse Event Malfunction Summary report: N

ZIMMER NATURAL NAIL CM LAG SCREW REAMER

MDR report key: 3172106 · Received June 11, 2013

Report

Report Number
1822565-2013-00946
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 9, 2013
Report Date
May 13, 2013
Manufacturer
ZIMMER, INC.
Product Code
HSB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PART HAD A POTENTIAL FIELD AGE OF 6 MONTHS AND APPEARED TO HAVE BEEN HEAVILY USED, WITH SCUFFS ALONG THE REAMING TIP. IT IS POSSIBLE THAT THE REAMER CONTACTED THE NAIL DURING DRILLING, CAUSING THE TIP TO FRACTURE. THERE IS INSUFFICIENT INFORMATION TO DETERMINE CAUSE OF FRACTURE. DEVICE HISTORY RECORDS INDICATE ALL COMPONENTS WERE MANUFACTURED AND INSPECTED TO SPECIFICATION. THE PART WAS RETURNED FOR INSPECTION AND A FRACTURE WAS OBSERVED IN THE CUTTING TIP OF THE REAMER. AFTER MEASURING THE PART, NO DEVIATIONS FROM SPECIFICATIONS WERE FOUND INT HE LOCATIONS MEASURED.

Description of Event or Problem · 1

IT IS REPORTED THAT WHILE REAMING FOR THE LAG SCREW THE TIP BROKE OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264776 ZIMMER NATURAL NAIL CM LAG SCREW REAMER HSB ZIMMER, INC. 62211917

Patients

Seq Age Sex Outcome Treatment