PLUM A+ DRIVER ED 2
Report
- Report Number
- 9615050-2013-01588
- Event Type
- Malfunction
- Date Received
- June 11, 2013
- Date of Event
- May 15, 2013
- Report Date
- May 15, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
TESTING AND INVESTIGATION FOUND THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. THIS WAS DUE TO THE PIEZO ALARM ASSEMBLY. FURTHER TESTING BY THE SUPPLIER FOUND THAT THE TRANSISTOR GAIN VALUE (BETA) OF THE TRANSISTOR, INTERNAL TO THE PIEZO, WAS NOT SUFFICIENT TO DRIVE THE PIEZO BUZZER, RESULTING IN NO AUDIBLE ALARM TONE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "E301 - AUDIO ALARM FAILURE." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 264815 | PLUM A+ DRIVER ED 2 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PLUM A+ SOFTWARE MODULE, LIST #20791| SN (B)(4) |