FDA Adverse Event Malfunction Summary report: N

PLUM A+ DRIVER ED 2

MDR report key: 3172082 · Received June 11, 2013

Report

Report Number
9615050-2013-01588
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 15, 2013
Report Date
May 15, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K042081
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. THIS WAS DUE TO THE PIEZO ALARM ASSEMBLY. FURTHER TESTING BY THE SUPPLIER FOUND THAT THE TRANSISTOR GAIN VALUE (BETA) OF THE TRANSISTOR, INTERNAL TO THE PIEZO, WAS NOT SUFFICIENT TO DRIVE THE PIEZO BUZZER, RESULTING IN NO AUDIBLE ALARM TONE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "E301 - AUDIO ALARM FAILURE." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264815 PLUM A+ DRIVER ED 2 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PLUM A+ SOFTWARE MODULE, LIST #20791| SN (B)(4)