FDA Adverse Event Malfunction Summary report: N

PLUMA+ UK WITH 3PIN

MDR report key: 3172081 · Received June 11, 2013

Report

Report Number
9615050-2013-01587
Event Type
Malfunction
Date Received
June 11, 2013
Date of Event
May 1, 2013
Report Date
May 13, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
Z-0624-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE DOOR ROLLER AND DOOR ROLLER PIN WERE BROKEN OFF. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE DOOR ROLLER WAS BROKEN. THE DEVICE WAS RETURNED TO THE CLINICAL ENGINEERING DEPARTMENT FOR A REPORT OF, FAULTY. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING OR, EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
264355 PLUMA+ UK WITH 3PIN 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PLUM A+ SOFTWARE MODULE: LIST #12097,| SN (B)(4)