FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3172040 · Received June 10, 2013

Report

Report Number
1720753-2013-06884
Event Type
Malfunction
Date Received
June 10, 2013
Date of Event
May 29, 2013
Report Date
June 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE HARD DRIVE WAS REPLACED, THE SYSTEM SOFTWARE WAS RELOADED, AND THE CINE DRIVE WAS REFORMATTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM LOCKED UP DURING A PROCEDURE. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
261590 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1