FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3172039 · Received June 10, 2013

Report

Report Number
1720753-2013-06886
Event Type
Malfunction
Date Received
June 10, 2013
Date of Event
May 29, 2013
Report Date
June 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE BATTERY PACK WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A "KV ON IN ERROR" ERROR MESSAGE. THIS ERROR MESSAGE LIKELY CAUSED THE SYSTEM TO SHUT DOWN. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
261663 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1