FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3172008 · Received June 10, 2013

Report

Report Number
1720753-2013-06914
Event Type
Malfunction
Date Received
June 10, 2013
Date of Event
May 29, 2013
Report Date
June 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SINGLE BOARD COMPUTER WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM HAD VERTICAL LINES ON THE MONITOR. THE FSE NOTED THE SYSTEM WOULD NO LONGER BOOT UP AND THE DEBUG MONITOR SHOWED A FATAL FRAME SYNC ERROR. THERE IS NO REPORT OF PT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
261653 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1