PLUM A+ SOFTWARE MODULE, LIST #20791, SN UNK
Report
- Report Number
- 9615050-2013-01578
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- January 1, 2013
- Report Date
- May 22, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- Z-0624-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. AS INDICATED, THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE DOOR ROLLER WAS BROKEN OFF. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 261648 | PLUM A+ SOFTWARE MODULE, LIST #20791, SN UNK | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PLUM A+ SOFTWARE MODULE, LIST #20791, SN UNK |