GEMSTAR 7 THERAPY
Report
- Report Number
- 9615050-2013-01571
- Event Type
- Malfunction
- Date Received
- June 10, 2013
- Date of Event
- May 14, 2013
- Report Date
- May 14, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K083019
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT ALARM WHEN A PROXIMAL OCCLUSION WAS PRESENT. THIS IS DUE TO CONTAMINATION ON THE PROXIMAL PRESSURE SENSOR PIN. THE CONTAMINATION PREVENTED THE PROXIMAL PRESSURE SENSOR PIN FROM CORRECTLY CONTACTING THE ADMINISTRATION SET CASSETTE DIAPHRAGM; THEREFORE, THE PUMP DID NOT DETECT PRESSURE CHANGES IN THE PROXIMAL TUBING. THE CONTAMINATION IS A RESULT OF USE IN THE CUSTOMER ENVIRONMENT. THIS REPORT REPRESENTS ALL THE INFO KNOW BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED DURING PREVENTATIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE DID NOT ALARM WHEN A PROXIMAL OCCLUSION WAS PRESENT. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.