FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3171988 · Received June 10, 2013

Report

Report Number
1720753-2013-06885
Event Type
Malfunction
Date Received
June 10, 2013
Date of Event
May 29, 2013
Report Date
June 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE VIDEO CABLES WERE RESEATED, AND THE FLUORO FUNCTIONS BOARD VOLTAGES WERE ADJUSTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT DISPLAY A LIVE FLUOROSCOPY IMAGE, THUS MAKING THE SYSTEM UNUSABLE. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
261475 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1