FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3171938 · Received June 7, 2013

Report

Report Number
1720753-2013-06855
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
May 28, 2013
Report Date
June 7, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE MAIN POWER PLUG TERMINAL WAS RETIGHTENED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WAS SHUTTING DOWN IN THE MIDDLE OF SCREENING. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
253711 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1