FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3171935 · Received June 7, 2013

Report

Report Number
1720753-2013-06859
Event Type
Malfunction
Date Received
June 7, 2013
Date of Event
May 29, 2013
Report Date
June 7, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE CONTROL PANEL PROCESSOR AND CONTROL PANEL I/O BOARDS WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED COMMUNICATION AND CONTROL PANEL ERROR MESSAGES. THESE ERRORS ARE LIKELY TO RESULT IN A SYSTEM LOCK UP OR PREVENT THE SYSTEM FROM BOOTING TO A USABLE STATE. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
254036 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1