Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE CONTROL PANEL PROCESSOR AND CONTROL PANEL I/O BOARDS WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM DISPLAYED COMMUNICATION AND CONTROL PANEL ERROR MESSAGES. THESE ERRORS ARE LIKELY TO RESULT IN A SYSTEM LOCK UP OR PREVENT THE SYSTEM FROM BOOTING TO A USABLE STATE. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.