FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ SINGLE CHANNE
MDR report key: 3171889
·
Received June 7, 2013
Report
- Report Number
- 9615050-2013-01553
- Event Type
- Malfunction
- Date Received
- June 7, 2013
- Date of Event
- May 16, 2013
- Report Date
- May 16, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K041550
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND AT POWER UP, THE DEVICE DISPLAYED A WHITESCREEN ERROR AND THE DEVICE SOUNDED AN AUDIBLE ALARM FROM THE SECONDARY BEEPER. THIS WAS DUE TO THE USER INTERFACE CONTROLLER 2 (UIC2) BBRAM HARDWARE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A WHITESCREEN ERROR. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTS DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 253766 | SYMBIQ SINGLE CHANNE | 8FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |