FDA Adverse Event Malfunction Summary report: N

SONOSURG SCISSORS 5MM O.D., HF SERIES

MDR report key: 3171861 · Received May 23, 2013

Report

Report Number
8010047-2013-00120
Event Type
Malfunction
Date Received
May 23, 2013
Date of Event
April 25, 2013
Report Date
April 26, 2013
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
LFL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PROBE BROKE DOWN AT 15.5 MM FROM THE DISTAL END. THERE WAS NO SCRATCH OR CRACK FOUND IN THE FRACTURES SURFACE. THE SHAPE OF THE FRACTURE SURFACE SHOWED THAT THE FRACTURE DEVELOPED FROM THE SIDE SURFACE OF THE PROBE. THERE WAS NO SIGNIFICANT WEAR OF PTFE PAD. THE MFG HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. AS THE RESULT OF EVALUATION, IT IS HIGHLY LIKELY THAT THE PHYSICIAN ACTIVATED THE ULTRASONIC OUTPUT WHILE THE SUBJECT DEVICE WAS TWISTED AND THE STRESS WAS PUT ON THE PROBE, RESULTED IN BREAK. THE INSTRUCTION MANUAL OF SONOSURG SCISSORS PROHIBITS THE USAGE AS MENTIONED IN THE ABOVE, AND DESCRIBES HOW TO DEAL WITH WHEN AN IRREGULARITY IS FOUND ON THE PROBE IS BROKEN. BASED UPON THE FINDING OF THE EVALUATION, THIS REPORT APPEARS TO BE RELATED TO USER HANDLING. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

WHEN A PHYSICIAN USED THE SUBJECT DEVICE FOR A LAPAROSCOPIC MYOMECTOMY, THE PROBE BROKE AND FELL OFF INTO THE PT'S BODY. THE BROKEN PIECE OF THE PROBE WAS TAKEN OUT. THE PHYSICIAN REPLACED THE SUBJECT DEVICE WITH A SIMILAR DEVICE. THE PROCEDURE WAS COMPLETED AS SCHEDULED. THERE WAS NO PT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
228205 SONOSURG SCISSORS 5MM O.D., HF SERIES SONOSURG SCISSORS LFL OLYMPUS MEDICAL SYSTEMS CORPORATION T3905 NA

Patients

Seq Age Sex Outcome Treatment
1