SCRDRIVERSHAFT T25 STRAIGHT TIP W/HEX-CO
MDR report key: 3171578
·
Received June 15, 2013
Report
- Report Number
- 8030965-2013-03087
- Event Type
- Malfunction
- Date Received
- June 15, 2013
- Report Date
- September 23, 2011
- Manufacturer
- SYNTHES GMBH
- Product Code
- HXX
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(6). THE CHECKING OF MANUFACTURING DATA AND HARDNESS SPECIFICATIONS ALSO SHOWED THAT THERE WERE NO DEVIATIONS FROM THE SPECIFICATIONS. INVESTIGATION OF THE AFFECTED SCREWDRIVER HAS SHOWN THAT THE TIP WAS COMPLETELY BROKEN OFF BY EXCESSIVE TORSION.
Description of Event or Problem · 1
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: THE TIP OF THE INSTRUMENT BROKE OFF ON AN UNKNOWN DATE. THE BROKEN PARTS WERE RETRIEVED. IT IS UNKNOWN WHERE THIS OCCURRED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 1 OF 1 FOR (B)(4).