SCRDRIVERSHAFT T25 F/URS
Report
- Report Number
- 8030965-2013-03397
- Event Type
- Malfunction
- Date Received
- June 15, 2013
- Date of Event
- September 14, 2011
- Report Date
- September 15, 2011
- Manufacturer
- SYNTHES GMBH
- Product Code
- HXX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. THE DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES, BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. THE DEVICE WAS RETURNED FOR INSPECTION AND THE EVENT WAS CONFIRMED. THE SCREWDRIVER SHATTERED DURING LOCKING AND IT IS CLEARLY VISIBLE THAT THE TIP WAS TWISTED BEFORE THE FINAL BREAKAGE. THE BROKEN SURFACE IS HOMOGENOUS WHICH INDICATES MATERIAL CONFORMITY. BASED ON THESE FINDINGS WE CONCLUDE THAT THE CAUSE OF THE FAILURE IS NOT DUE TO ANY MANUFACTURING NON CONFORMANCES. THESE FINDINGS ARE INDICATIONS THAT TOO MUCH TORQUE WAS APPLIED ONTO THE INSTRUMENT WHILE TIGHTENING THE SCREW. A REVIEW OF THE DHR INDICATES THERE WERE NO ANOMALIES WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THIS COMPLAINT. THE MANUFACTURING RECORDS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. ALL RECORDS INDICATE THE PRODUCT WAS MANUFACTURED TO SPECIFICATIONS. NO PRODUCT FAULT COULD BE DETECTED. CONCLUSION ¿ THE COMPLAINT IS CONSIDERED INDETERMINATE FROM A MANUFACTURING PERSPECTIVE. THE INVESTIGATION RESULTS CONCLUDED THAT TOO MUCH TORQUE WAS APPLIED ONTO THE INSTRUMENT WHILE TIGHTENING THE SCREW. THUS, THE DEVICE FAILURE IS RELATED TO THE CONDITIONS OF USE AND THE OPERATIONAL CONTEXT CONTRIBUTED TO THIS EVENT. PLEASE NOTE THAT OUR PROJECT MANAGER HAS DECIDED TO TAKE A CLOSER LOOK AT THE DESIGN IN ORDER TO AVOID PROBLEMS LIKE THIS IN THE FUTURE.
IT WAS REPORTED THAT THE SCREWDRIVER WAS FOUND TWISTED AT THE END. THE SURGEONS WERE LOCKING DOWN THE CAPS, USING THE CORRECT TECHNIQUE, WITH A COUNTER TORQUE AND TORQUE LIMITING HANDLE WHEN THE SCREWDRIVER SHATTERED DURING LOCKING. THERE WAS NO INJURY TO THE PATIENT AND ALL THE PIECES OF THE DRIVER WERE LOCATED AND REMOVED. NO FURTHER INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272581 | SCRDRIVERSHAFT T25 F/URS | HXX | SYNTHES GMBH | 3680245 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |